Chapter 3: Testing Performance

To test the single-clock, single-edge, state acquisition

cOn the oscilloscope, select [Shift] Time, then select [Enter] to display the setup time (Time(1)-(2)).

dAdjust the pulse generator channel 1 Delay until the pulses are aligned according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

3Select the logic analyzer sample positions:

a Click the Sampling Setup icon. The Analyzer Setup dialog opens with the Sampling tab displayed.

b Click Sample Positions....

c In the Sample Positions dialog, drag the blue bar for My Bus 1 to the sample position of the first setup/hold combination to be tested (or enter the value in the signal fields).

42

Page 42
Image 42
Agilent Technologies 1680, 1690 manual To test the single-clock, single-edge, state acquisition