Agilent Technologies 1680, 1690 manual To test disk drive voltages

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 90
Image 90
To test disk drive voltages

Chapter 5: Troubleshooting

Troubleshooting the Agilent 1680A,AD-series

4Using an oscilloscope, probe pins 39 and 40 of J111 for +3.3Vdc

If +3.3Vdc is present on J111 of pins 39 and 40, and digital signals are present on the video data pins indicated above, then the CPU board video circuit is operating properly.

5Remove power. Allow time for the capacitors in the power supply to discharge before disconnecting the power supply, doing the repair, and reassembling the instrument.

To test disk drive voltages

The following procedure is a guide to help further identify possible problems with either the flexible disk drive or hard disk drive.

Equipment Required

Equipment

Critical Specification

Recommended Model/Part

Digitizing Oscilloscope

> 100 MHz Bandwidth

54600B

1Close the Agilent Logic Analyzer application and all other applications running on the logic analyzer.

2Shut down the instrument.

a Click on the Start button in the task bar, then select Shut Down.

b In the Shut Down window, select Shut Down from the menu, then select the OK button.

3After the instrument turns off, unplug the instrument and remove the cover.

4Disconnect the suspect disk drive. Remove the disk drive from the chassis and reconnect the cable.

5Troubleshoot a hard disk drive.

a Apply power to the instrument.

90

Page 90
Image 90
Agilent Technologies 1680, 1690 manual To test disk drive voltages