Chapter 6: Replacing Assemblies

1680A,AD-series disassembly/assembly

To remove the power supply

1Do the procedure "To remove the chassis from the sleeve".

2Disconnect the 16-pin power supply cable and the power sense cable from the distribution board.

3Using a Torx T-15 screwdriver, remove four screws that secure the power supply to the chassis.

4Lift the power supply out of the chassis.

5Disconnect the 14-pin and the 10-pin power supply cables from the distribution board.