Chapter 5: Troubleshooting

Troubleshooting the Agilent 1690A,AD-series

3A

Uninstall, then reinstall

the Agilent Logic

Analyzer application.

Does the

No

 

 

 

application

 

 

 

 

 

 

 

 

software launch?

 

 

 

 

 

 

Yes

 

 

Possible problem with

 

 

 

 

 

 

 

Do error

Yes

host operating system.

messages appear?

 

 

 

 

Consult host PC

 

 

 

 

 

 

 

 

documentation.

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

3B

Is the IEEE 1394

No

Reconnect or reseat

cable connected and

 

 

 

 

IEEE 1394 cable.

seated?

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Perform " To

 

 

 

verify connectivity"

 

 

 

in Chapter 5.

 

 

 

 

 

 

 

 

 

 

 

 

 

Is connectivity

No

Uninstall, then reinstall

the Agilent Logic

 

 

 

 

verified?

 

 

Analyzer application.

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

Remove the cover and observe the status LEDs on the acquisition board.

Are any green

Yes

 

LEDs blinking?

 

No

Possible problem with PCI IEEE 1394 board on host PC. Replace

the board.

Possible problem with

acquisition board.

Replace acquisition bd.