Agilent Technologies 1690, 1680 manual Connect and configure the logic analyzer

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 67
Image 67
3Set up the function generator according to the following table.

Chapter 3: Testing Performance

To test the time interval accuracy

3Set up the function generator according to the following table.

Function Generator Setup

Freq:

40.000 MHz

Ampl:

1.00 Vpp

Offset:

0.0 mV

Modulation

Off

 

 

Connect and configure the logic analyzer

1Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse generator channel 1 output.

2Using the SMA cable and the BNC adapter, connect the External Input of the pulse generator to the Main Signal of the function generator.

3Enable the function generator output and the pulse generator Channel 1 output.

4Configure the Analyzer Setup dialog: a Click the Manual background Sampling Setup icon.

b In the Analyzer Setup dialog, select Timing - Asynchronous Sampling. c Configure Trigger Position - 100% poststore.

67

Page 67
Image 67
Agilent Technologies 1690, 1680 manual Connect and configure the logic analyzer