Agilent Technologies 1680, 1690 manual To remove the distribution board

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 138
Image 138
Manual background

Chapter 6: Replacing Assemblies

1690A,AD-series disassembly/assembly

5Lift the power supply out of the chassis.

6Reverse this procedure to install the power supply

When installing a replacement power supply, transfer both the power sense cable and the power supply on/off cable to the replacement power supply.

To remove the distribution board

1Do the following procedures:

To remove the chassis from the sleeve

To remove the front panel assembly

To remove the deck

2Disconnect the following cables from the distribution board:

power supply output cables (P1, P2)

on/off cable (P3)

power sense cable (P7)

fan cables (P5, P6)

3Using a Torx T-10 screwdriver, remove three screws that secure the distribution board to the chassis.

138

Page 138
Image 138
Agilent Technologies 1680, 1690 manual To remove the distribution board