Chapter 3: Testing Performance

To set up the logic analyzer for the state mode tests

3Configure the trigger according to your logic analyzer:

aIn the Listing window, click on the trigger pattern field for My Bus 1 to select.

bEnter the following pattern for your logic analyzer.

1680A,AD, 1690A,AD - "AA" 1681A,AD, 1691A,AD - "2A" 1682A,AD, 1692A,AD - "AA" 1683A,AD, 1693A,AD - "A"

cClick the Trigger Setup icon.

dFor the Default Storage, select Store Nothing.

eClick OK to close the Advanced Trigger dialog.

4Activate the pulse generator data and clock outputs:

a On the pulse generator, enable the channel 1 OUTPUT, channel 1 OUTPUT, channel 2 OUTPUT and channel 2 OUTPUT (LEDs off)

b On the pulse generator, enable the trigger OUTPUT. (LED off)

5Set up the Markers:

The following procedure is done after the first run of test data is acquired (during one of the state clock mode tests).

a From the main menu, choose Markers>Properties....

b In the Marker Properties tab of the Listing Viewer Properties dialog,

34