3Select the clocks to be tested:

Chapter 3: Testing Performance

To test the multiple-clock state acquisition

bIn the oscilloscope timebase menu, select Position. Using the oscilloscope knob, position the data waveform so the falling edge is near the center of the display.

cOn the oscilloscope, select [Shift] Time, then select [Enter] to display the setup time (Time(1)-(2)).

dAdjust the pulse generator channel 1 Delay until the pulses are aligned according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

3Select the clocks to be tested:

a Click the Manual background Sampling Setup icon. The Analyzer Setup dialog opens.

b In the Sampling tab, click the Master button for one of the clocks and select Rising Edge.

c Repeat the above steps for each of the remaining clocks until all clocks have been configured with Rising Edge.

53

Page 53
Image 53
Agilent Technologies 1690, 1680 manual Select the clocks to be tested