Chapter 3: Testing Performance

To test the single-clock, multiple-edge, state acquisition

To test the single-clock, multiple-edge, state acquisition

Testing the single-clock, multiple-edge, state acquisition verifies the performance of the following specifications:

Minimum master-to-master clock time.

Maximum state acquisition speed.

Setup/Hold time for single-clock, multiple-edge, state acquisition.

This test checks two combinations of data using a multiple-edge single clock at two selected setup/hold times.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

 

 

 

Pulse Generator

200 MHz 3.0 ns pulse width, < 600 ps rise time

8133A option 003

Digitizing Oscilloscope

6 GHz bandwidth, < 58 ps rise time

54750A w/ 54751A

Adapter

SMA(m)-BNC(f)

1250-1200

SMA Coax Cable (Qty 3)

18 GHz bandwidth

8120-4948

Coupler

BNC(m)(m)

1250-0216

BNC Test Connector,

 

 

6x2 (Qty 4)

 

 

 

 

 

Set up the equipment

1If you have not already done so, do the following procedures:

“To set up the test equipment and the logic analyzer” on page 23 “To set up the logic analyzer for the state mode tests” on page 33

2Modify the following pulse generator settings:

Period: 10.000 ns

Channel 2: Width 3.000 ns

Channel 2: Pulse 1

Channel 1: Pulse

Channel 1: Width 5.000 ns

57

Page 57
Image 57
Agilent Technologies 1690, 1680 manual To test the single-clock, multiple-edge, state acquisition