Chapter 5: Troubleshooting
General Troubleshooting
To test the auxiliary power
The +5 V auxiliary power is protected by a current overload protection device. If the current on pins 1 and 39 exceed 0.33 amps, the circuit will open. When the short is removed, the circuit will reset in approximately 1 minute. There should be +5 V after the 1 minute reset time.
Equipment Required
Equipment | Critical Specifications | Recommended | |
Model/Part | |||
|
| ||
Digital Multimeter | 0.1 mV resolution, better than E2373A | ||
| 0.005% accuracy |
|
•Use the multimeter to verify the +5 V on pins 1 and 39 of the probe cables.