Agilent Technologies 1690, 1680 manual Check the setup/hold combination

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 41
Image 41
Check the setup/hold combination

Chapter 3: Testing Performance

To test the single-clock, single-edge, state acquisition

cIf the pulse width is outside the limits, adjust the pulse generator channel 2 width until the pulse width is within limits.

Check the setup/hold combination

The following setup/hold combinations will be tested:

Setup/Hold Combinations

Test

Setup/Hold

Setup/Hold

Sample Position

Combination

Times

Window

(in middle of Window)

 

 

 

 

1

4.50/-2.0 ns

2.5 ns

-3.25 ns

2

-2.0/4.50 ns

2.5 ns

+3.25 ns

1Disable the pulse generator channel 1 COMP (with the LED off).

2Using the Delay mode of the pulse generator channel 1, position the pulses

according to the setup time of the setup/hold combination selected,

+0.0 ps or 100 ps as measured on the oscilloscope:

a On the Oscilloscope, select [Define meas] Define Time - Stop edge: rising, Edge number 2.

b In the oscilloscope timebase menu, select Position. Using the oscilloscope knob, position the data waveform so the falling edge is near the center of the display.

41

Page 41
Image 41
Agilent Technologies 1690, 1680 manual Check the setup/hold combination