Chapter 5: Troubleshooting
Troubleshooting the Agilent
2
Are the power | No | Ensure the power cord |
|
| |||
supply fans |
|
|
|
| |||
|
| is properly connected. |
|
| |||
running? |
|
|
|
| |||
|
|
|
|
|
| ||
| Yes |
|
|
|
| No | |
|
|
|
|
| |||
|
|
|
|
| |||
|
|
|
| Is the power cord | |||
|
|
|
|
|
| ||
|
|
|
| connected? |
|
| |
|
|
|
|
| Yes |
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
| |
|
| No | Suspect the power supply. | ||||
Are the instrument | Perform "To test the |
|
| ||||
|
|
|
| ||||
fans running? |
|
| power supply voltages" |
|
| ||
|
|
|
| in Chapter 5. |
|
| |
| Yes |
|
|
|
| No | |
|
|
|
|
| |||
|
|
|
| Do all test | |||
|
|
|
|
|
| ||
|
|
|
| points pass? |
|
| |
|
|
|
|
| Yes |
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| Ensure all instrument |
|
| |
|
|
|
| cables are connected |
|
| |
|
|
|
| and properly seated. |
|
| |
|
|
|
|
|
|
|
|
Reconnect the
power cord.
Replace the
power supply.
1
96