Agilent Technologies 1680, 1690 manual Power Requirements, Operating Environment, Storage

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 16
Image 16
Power Requirements

Chapter 2: Preparing for Use

Power Requirements

The logic analyzer requires a power source of either 115 Vac or 230 Vac, –22 % to +10 %, single phase, 48 to 66 Hz, CAT II pollution degree 2, 140/400 Watts nominal maximum power (1680A/AD-series), and 76/200 Watts nominal maximum power (1690A/AD-series).

Operating Environment

The operating environment is listed in chapter 1. The logic analyzer will operate at all specifications within the temperature and humidity range given in chapter

1.However, reliability is enhanced when operating the logic analyzer within the following ranges:

Temperature: +20° C to +35° C (+68° F to +95° F)

Humidity: 20% to 80% noncondensing

Note the recommended noncondensing humidity. Condensation within the instrument can cause poor operation or malfunction. Provide protection against internal condensation.

Storage

Store or ship the logic analyzer in environments within the following limits:

Temperature: -40°C to +75° C

Humidity: Up to 90% at 65° C

Altitude: Up to 15,300 meters (50,000 feet)

Protect the logic analyzer from temperature extremes which cause condensation on the instrument.

To inspect the logic analyzer

1Inspect the shipping container for damage.

If the shipping container or cushioning material is damaged, keep them until you have checked the contents of the shipment and checked the instrument mechanically and electrically.

2Check the supplied accessories.

Accessories supplied with the logic analyzer are listed in “Accessories” on page 10.

16

Page 16
Image 16
Agilent Technologies 1680, 1690 manual Power Requirements, Operating Environment, Storage, To inspect the logic analyzer