Chapter 3: Testing Performance

To test the time interval accuracy

To test the time interval accuracy

Testing the time interval accuracy does not check a specification, but does check the following:

125 MHz oscillator

This test verifies that the 125 MHz timing acquisition synchronizing oscillator is operating within limits.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Pulse Generator

Function Generator

SMA Cable

Adapter

BNC Test Connector, 6x2

200 MHz 2.5 ns pulse width, < 600 ps rise time Accuracy(5)(10-6)x frequency

BNC(m)-SMA(f)

8133A option 003

33250A

8120-4948

1250-2015

Set up the equipment

1Set up the logic analyzer:

a If you have not already done so, do the procedure “To set up the test equipment and the logic analyzer” on page 23.

b Exit and restart the Agilent Logic Analyzer applications to reinitialize the logic analyzer.

2Set up the pulse generator according to the following table.

Pulse Generator Setup

Timebase

Channel 1

Trigger

Mode: Ext

Mode: Square

Divide: Divide ⎟ 1

 

Delay: 0.000 ns

Ampl: 0.50 V

 

High: -0.90 V

Ampl: 0.50 V

 

Low: -1.70 V

Offs: 0.00V

 

COMP: Disabled (LED off)

 

 

 

 