Test the 0 V User Threshold

Chapter 3: Testing Performance

To test the threshold accuracy

Test the 0 V User Threshold

1Set up the logic analyzer:

a On the logic analyzer, click the Manual background Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b In the Analyzer Setup dialog, click the threshold field for Pod 1. The Threshold Settings dialog appears.

c In the Threshold Settings dialog, select User Defined and enter 0 V in the associated field.

dClick OK to close the Threshold Settings dialog.

eClick OK to close the Analyzer Setup dialog.

2Test the high-to-low transition:

a On the DC source, enter a voltage setting of –0.064 V.

b On the logic analyzer, click the Manual background Run icon. The display should show all channels at a logic "0".

3Test the low-to-high transition:

a On the DC source, enter a voltage setting of 0.064 V.

b On the logic analyzer, click the Manual background Run icon. The display should show all channels at a logic "1" (0x1FFFF).

4Record a PASS/FAIL in the performance test record for Threshold Accuracy Pod 1 - User 0 V.