To set up the logic analyzer for the state mode tests

Chapter 3: Testing Performance

To set up the logic analyzer for the state mode tests

To set up the logic analyzer for the state mode tests

1Set up the logic analyzer:

a If you have not already done so, do the procedure “To set up the test equipment and the logic analyzer” on page 23.

b Exit and restart the Agilent Logic Analyzer application to reinitialize the logic analyzer.

2Configure the Analyzer Setup dialog: a Click the Manual background Sampling Setup icon.

b Select State - Synchronous Sampling.

c Configure Trigger Position - 100% poststore. d Select an Acquisition Depth of 8K.

eClick the Buses/Signals tab.

fClick the threshold field of one of the pods. The Threshold Settings dialog appears.

gIn the Threshold Settings dialog, select Standard and ECL (–1.30 V).

hClick OK to close the Threshold Settings dialog.

iClick OK to close the the Analyzer Setup dialog.