Chapter 5: Troubleshooting

Troubleshooting the Agilent 1690A,AD-series

aClick on the Start button in the task bar, then select Shut Down.

bIn the Shut Down window, select Shut Down from the menu, then select the OK button.

3Remove the power supply from the instrument. Refer to “To remove the power supply” in Chapter 6.

4After removing the poser supply, connect a power cord to the power supply and plug the power cord into line power.

5Using DVM, measure the power supply voltages.

Power Supply Voltages

CN1

 

CN2

 

 

 

 

 

Pin

Voltage

Pin

Voltage

 

 

 

 

1-3

+3.4 V

1,2

+5.1 V

4,5

COM

3

COM

6,7

+3.4 V

4

-5.2 V

8-10

COM

5,6

COM

 

 

7

+12 V

 

 

8

+5.1 V

 

 

9,10

COM

 

 

11,12

-5.2 V

 

 

13

COM

 

 

14

+12 V

 

 

 

 