Chapter 6: Replacing Assemblies

1690A,AD-series disassembly/assembly

1690A,AD-series disassembly/assembly

Prepare the instrument for disassembly

Do this procedure before doing any disassembly procedure on the instrument.

1Close the Agilent Logic Analyzer application software.

2Disconnect the logic analyzer from the host PC.

3Remove power and disconnect the power cord.

4Move the instrument to a static safe work environment.

To remove the chassis from the sleeve

Before disassembling the instrument, it must be turned off and placed in a static safe work environment. If you haven't already done so, do the previous procedure "Prepare the instrument for disassembly."

1Using a Torx T-15 screwdriver, remove the two screws that secure the handle to the side of the instrument and lift off the handle.

2Using a Torx T-10 screwdriver, remove five screws that secure the logic analyzer cables (and spacers, if installed) to the front panel of the logic analyzer.

3Disconnect the logic analyzer cables from the front panel. Remove the logic analyzer cables (and spacers, if installed) from the logic analyzer.

4Using a Torx T-10 screwdriver, remove thirteen screws that secure the cover to the chassis.

5With the logic analyzer upright, slide the chassis out of the cover.