To verify the CD-ROM

Chapter 5: Troubleshooting

Troubleshooting the Agilent 1680A,AD-series

eWhile the instrument is attempting to save the file to flexible disk, probe for digital signals on the flexible disk drive connector according to the following table.

Disk Drive Voltages

Pin

Signals

Pin No.

Signal

1

+5 V

2

INDEX

3

+5 V

4

DRIVE SELECT

5

+5 V

6

DISK CHANGE

7

NC

8

READY

9

NC

10

MOTOR ON

11

NC

12

DIRECTION SELECT

13

NC

14

STEP

15

0 V

16

WRITE DATA

17

0 V

18

WRITE GATE

19

0 V

20

TRACK 00

21

O V

22

WRITE PROTECT

23

0 V

24

READ DATA

25

0 V

26

SIDE ONE SELECT

 

 

 

 

7Repeat steps 1 through 3 above. After the instrument turns off, unplug the instrument.

8Replace suspect disk drive if necessary, then reassemble the instrument.

To verify the CD-ROM

The CD-ROM drive itself can be tested using an audio CD. Install CD player-style headphones in the CD-ROM audio output jack. With the instrument powered on, insert an audio CD into the CD-ROM drive. If the CD-ROM is operating properly, it should begin playing the audio CD.

92

Page 92
Image 92
Agilent Technologies 1680, 1690 manual To verify the CD-ROM