To test the LCD display signals

Chapter 5: Troubleshooting

Troubleshooting the Agilent 1680A,AD-series

To test the LCD display signals

Before attempting to do this procedure, ensure that the video signal cable connected to the PCI video board is properly seated. Attempt to reseat the cable two or three times. If other repairs were done to the instrument, and the video is now no longer operating, it is very likely that the video cable is not properly seated.

Refer to chapter 6, "Replacing Assemblies", for instructions to remove or replace covers and assemblies.

WARNING:Warning Hazardous voltages exist on the power supply and the LCD display, and the LCD inverter. This procedure is to be performed by service-trained personnel aware of the hazards involved, such as fire and electric shock.

1Remove the sleeve. Refer to chapter 6 for more information on how to remove the sleeve.

2Connect a power cord to the instrument and apply power.

3Using an oscilloscope, probe the following pins of J111 for digital signals.

2, 4, 5

9, 10, 11, 13, 14, 15

29, 30, 31, 33, 34, 35, 37

12

4140

01680f01.cdr

89

Page 89