Chapter 3: Testing Performance

Performance Test Record

Performance Test Record

 

Agilent 1680/90-Series Logic Analyzer_______

Serial No.______________________

Work Order No.___________________

Recommended Test Interval - 2 Years/4000 hours

Date___________________

Recommended next testing___________________

Temperature___________________

 

 

Test

Settings

Results

 

 

 

 

 

 

Self-Tests

 

 

Pass/Fail

________

 

 

 

 

Threshold

± (65 mV + 1.5% of threshold setting)

 

 

Accuracy

 

 

 

 

Pod 1

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

Pod 2

ECL, ±84mV

Pass/Fail

________

 

0

V, ± 65mV

Pass/Fail

________

Pod 3

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

Pod 4

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

Pod 5

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

Pod 6

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

Pod 7

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

Pod 8

ECL, ±84 mV

Pass/Fail

________

 

0

V, ±65 mV

Pass/Fail

________

 

 

 

 

 

71

Page 71
Image 71
Agilent Technologies 1690, 1680 manual Performance Test Record, Test Settings Results Self-Tests, Threshold, Accuracy