Agilent Technologies 1690 Troubleshoot a flexible disk drive. a Apply power to the instrument

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 91
Image 91
6Troubleshoot a flexible disk drive. a Apply power to the instrument.

Chapter 5: Troubleshooting

Troubleshooting the Agilent 1680A,AD-series

bAs the instrument is booting, probe for digital signals on the hard disk drive connector according to the following table.

Disk Drive Voltages

Pin No.

Signal

Voltage

Pin No.

Signal

Voltage

Pin No.

Signal

Voltage

1

RESET

 

27

IORDY

 

34

PDIAG

 

3 - 18

DATA

 

28

CSEL

 

35

DA00

 

20

KEY

 

29

DMACK

 

36

DA02

 

21

DMARQ

 

31

INTRQ

 

37

CS0

 

23

DIOW

 

32

IOCS16

 

38

CS1

 

25

DIOR

 

33

DA01

 

39

DASP

 

Pins 2, 19, 22, 24, 26, 30, 40, are GROUND

Pins 41 and 42 are +5 Vdc

6Troubleshoot a flexible disk drive. a Apply power to the instrument.

b After the instrument finished booting, launch the Agilent Logic Analyzer application.

c Insert a formatted flexible disk in the instrument flexible disk drive.

d Attempt to do a File Save of the Agilent Logic Analyzer default configuration to the flexible disk drive.

91

Page 91
Image 91
Agilent Technologies 1690, 1680 manual Troubleshoot a flexible disk drive. a Apply power to the instrument