Chapter 3: Testing Performance

To test the multiple-clock state acquisition

Connect the 1682/83/92/93A,AD Logic Analyzer to the Pulse Generator

Testing

Connect to 8133A

Connect to 8133A

Connect to 8133A

Combination

Channel 2 Output

Channel 2 Output

Channel 1 Output

1

Pod 1, channel 3

 

Pod 2, channel 3

 

Pod 3, channel 3

 

Pod 4, channel 3

*1682A, AD or 1692A, AD only.

Pod 1, channel 11 Pod 2, channel 11 Pod 3, channel 11 * Pod 4, channel 11 *

Clock/data channel for Pod 1, 2, 3, and 4 (Clk 1, Clk 2, Clk 3, Clk 4)

3Activate the data channels that are connected according to one of the previous tables:

a Click the Clock/data channel for Pod 1, 2, 3, and 4 (Clk 1, Clk 2, Clk 3, Clk 4) Bus/Signal Setup icon. The Analyzer Setup dialog opens.

b Under the Buses/Signals tab, click Delete All at the bottom of the dialog.

c Using the mouse, activate the data channels being tested. Assign channels to bus/signal name My Bus 1.

d Click OK to close the Analyzer Setup dialog.