Chapter 5: Troubleshooting

General Troubleshooting

2Set up the pulse generator according to the following table.

Pulse Generator Setup

Timebase

Channel 2

Trigger

Channel 1

Mode: Int

Mode: Pulse

Divide: Divide 1

Mode: Square

Period: 20.000 ns

Divide: Square 1

Ampl: 0.50 V

Delay: 0.000 ns

 

 

Offs: 0.00 V

 

 

Ampl: 0.80 V

 

Ampl: 0.80 V

 

Offs: -1.30 V

 

Offs: -1.30 V

 

COMP: Disabled (LED Off)

 

COMP: Disabled (LED Off)

 

 

 

 

3Using four 6-by-2 test connectors, four BNC Couplers, and four SMA (m) - BNC (f) Adapters, connect the logic analyzer to the pulse generator channel outputs (to make the test connectors, see chapter 3, "Testing Performance"):

a Connect the even-numbered channels of the lower byte of the pod

under test and Clk 1 to the pulse generator channel 1 OUTPUT.

b Connect the odd-numbered channels of the lower byte of the pod under test to the pulse generator channel 1 OUTPUT.

c Connect the even-numbered channels of the upper byte of the pod under test to the pulse generator channel 2 OUTPUT.

d Connect the odd-numbered channels of the upper byte of the pod under test to the pulse generator channel 2 OUTPUT.

4Configure the Analyzer Setup dialog:

a In the Waveform window, click on the Manual background Sampling Setup icon.

b In the Analyzer Setup dialog, select State - Synchronous Sampling. c Configure Trigger Position - 100% poststore.

d Select an Acquisition Depth of 8K.

108