Chapter 5: Troubleshooting

General Troubleshooting

General Troubleshooting

This section includes troubleshooting procedures that can be done on either the Agilent 1680A,AD- or 1690A,AD-series logic analyzers. Before any of these procedures can be done on an Agilent 1690A,AD-series logic analyzer, the logic analyzer must be connected to a host PC and both the host PC and the logic analyzer must be turned on.

To run the self-tests

The Self Test menu checks the major hardware functions of the logic analyzer to verify that it is working correctly. Self-tests can be performed all at once or one at a time. Refer to Chapter 8 for more information on the logic analyzer self-tests.

CAUTION:Because the most recently acquired data will be lost, be sure to save important data before running self tests.

1Do the following steps for an Agilent 1690A,AD-series:

a Connect the logic analyzer to a host PC and apply power.

b Apply power to the host PC and allow the PC to finish booting.

2For an Agilent 1680A,AD-series logic analyzer, apply power to the instrument and allow it to finish booting.

3Launch the Agilent Logic Analyzer application. When the application launches, observe the status field and ensure it reads “Online.”

4In the Agilent Logic Analyzer application, choose Help>Self Test... from the main menu.

CAUTION:If you have acquired data, a warning message appears, "Running self-tests will invalidate acquired data"; click OK to continue.

103

Page 103
Image 103
Agilent Technologies 1690, 1680 manual General Troubleshooting, To run the self-tests, 103