Chapter 5: Troubleshooting

General Troubleshooting

analyzer Trigger In BNC.

cUsing a BNC-banana cable, connect the voltmeter to the logic analyzer Trigger Out BNC.

The voltmeter will display a voltage approximately 3 Vdc.

4Configure the external trigger:

a Select the Manual background Trigger Setup icon.

b In the Advanced Trigger dialog, for Trigger Sequence Step 1, select “Arm in from” instead of “Anything” as the event to trigger on.

c Then, select “External trigger” as the source of the arming signal.

dClick OK to close the Advanced Trigger dialog.

5Verify the external trigger: a Select the Manual background Run icon.

The logic analyzer will report "Waiting in Trigger Step 1...". The voltmeter will display approximately 0 Vdc.

b On the DC source, enter a voltage setting of 3.000 V.

The voltmeter will display approximately 3 Vdc. The logic analyzer will trigger and display a waveform similar to the following:

cOn the DC source, re-enter the voltage setting of 0.000 V.

6Disconnect the test equipment from the logic analyzer.