Chapter 3: Testing Performance

To set up the test equipment and the logic analyzer

To set up the test equipment and the logic analyzer

Before testing the specifications of the Agilent 1680A,AD-series or 1690A,AD- series logic analyzer, the test equipment and the logic analyzer must be set up and configured.

These instructions include detailed steps for initially setting up the required test equipment and the logic analyzer. Before performing any or all of the tests in this chapter, the following steps must be done.

Equipment Required

Equipment

Critical Specifications

Recommended HP/Agilent

Model/Part

 

 

 

 

 

Pulse Generator

200 Mhz, 2.5 ns pulse width, <600 ps rise time

8133A option 003

Digitizing Oscilloscope

6 GHz bandwidth, < 58 ps rise time

54750A w/ 54751A

Digital Multimeter

0.1 mV resolution, 0.005% accuracy

3458A

Function Generator

DC offset voltage ±1.5 V

3325B Option 002

 

 

 

Set up the test equipment

1Turn on the required test equipment listed in the table above. Let them warm up for 30 minutes before beginning any test.

2Set up the pulse generator according to the following table.

Pulse Generator Setup

Timebase

Channel 2

Trigger

Channel 1

 

 

 

 

Mode: Int

Mode: Pulse

Divide: Divide 2

Mode: Square

Period: 5.000 ns

Divide: Pulse 2

Ampl: 0.50 V

Delay: 0.0 ps

 

Width: 2.500 ns

Offs: 0.00 V

Ampl: 0.80 V

 

Ampl: 0.80 V

 

Offs: -1.30 V

 

Offs: -1.30 V

 

COMP: Disabled

 

COMP: Disabled

 

(LED Off)

 

(LED Off)

 

 

 

 

 

 

3Set up the oscilloscope:

a Select Setup, then select Default Setup.

23

Page 23
Image 23
Agilent Technologies 1690, 1680 To set up the test equipment and the logic analyzer, Set up the test equipment, Model/Part