Chapter 5: Troubleshooting

Troubleshooting the Agilent 1680A,AD-series

6A

Uninstall, then reinstall

the Agilent Logic

Analyzer application.

Does the

No

 

 

 

application

 

 

 

 

 

 

 

 

software launch?

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Do error

Yes

Perform "To recover

messages appear?

 

 

the operating system"

 

 

 

 

 

 

in Chapter 5.

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

6B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Remove power from

 

 

 

 

 

instrument, remove

 

 

 

 

 

cover, check internal

 

 

 

 

 

IEEE 1394 cabling.

 

 

 

 

 

 

 

 

 

 

 

 

Are all cables

No

Reconnect or reseat

 

connected and

 

 

 

 

 

all internal cabling.

seated?

 

 

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reseat the PCI IEEE

 

 

* All PCI boards must be installed in

board in ISA slot 1*.

 

 

 

 

the correct slots. See "To remove

Apply power to the

 

 

 

 

the ISA board" in Chapter 6.

instrument.

 

 

 

 

 

 

 

 

 

No

Observe the status LEDs

 

 

Is the application

on the acquisition board.

 

 

Online?

 

 

Refer to Acquisition board

 

 

 

 

status LEDs.

 

Yes

 

 

 

 

 

 

 

 

 

 

 

Are any green

Yes

LEDs blinking?

 

No

Possible problem with PCI IEEE 1394 board. Replace the PCI IEEE 1394 board.

Possible problem with

acquisition board.

Replace acquisition bd.

1