Change the Magnification Adjustment
4.(Optional) Change the MAGNIFICATION ADJUSTMENT if desired by moving its slider or typing an integer value from
NOTES: If you change the magnification adjustment, the Scanner must refocus. As prompted by the message that occurs, remove a film holder if present and place the calibration mask into the Scanner as explained in an earlier step of this section. Follow additional prompts as they appear to: remove the calibration mask, replace the film holder (with your film in it), and wait for focusing to complete.
The magnification adjustment for each film size is maintained from work session to work session.