Message | Possible Cause(s) | Suggested Solution(s) |
|
|
|
“Parity error detected | A hardware error has |
during transfer.” | been detected in the |
| cabling or in the |
| Scanner. |
Verify that all SCSI connections are correct (if necessary turn off all equipment to alter the connections). Try other cables if available. If the problem persists, contact Kodak.
“PLEASE WAIT . . . | This message appears | Wait for calibration to finish |
CALIBRATION IN | normally during | before inserting a Holder into the |
PROGRESS | calibration. | Scanner. |
DO NOT INSERT holders into Scanner.
WAIT for calibration to complete!”
“PLEASE WAIT . . . | Scanner is focusing. | As prompted, wait for focusing to |
FOCUSING IN |
| occur. |
PROGRESS” |
|
|
“Preferences file could not | The system disk is full, |
be saved.” | or is write protected (for |
| example on a SyQuest |
| disk), or there is a |
| problem with the system |
| disk. |
Create room on the system disk, remove write protection from the system disk, or have the system disk repaired.
“Problem with | A hardware problem was | Shut down the computer and the |
Magnification hardware.” | detected. | Scanner. Restart and try again. If |
|
| the problem persists, reinstall the |
|
| Scanner software as described in |
|
| Chapter 3 (Macintosh) or Chapter |
|
| 7 (PC), and try again. If the |
|
| problem continues contact |
|
| Kodak. |