Chapter 11. CALIBRATION

While pressing FUNC , press ENT and simultaneously.

& ENTER

& ENTER

PROG

SEG

0.0.0.0. Function test

0 0

0 0

*A d J S

SETUP

PROG

SEG

 

 

0 0

0 1

Key ON/OFF

Key test

 

 

Key ON/OFF

 

 

 

SETUP

 

PROG

SEG

 

 

 

 

ON/OFF of display

Display test

 

 

ON/OFF of display

 

 

 

SETUP

 

PROG

SEG

 

 

0 0

0 3

DI ON/OFF

DI test

 

 

SETUP

 

PROG

SEG

Input of Control

 

0 0

0 4

Control output test

output ON/OFF

 

 

 

Relay Voltage pulse

 

 

SETUP

 

PROG

SEG

Input of DO

 

0 0

0 5

Do output test

ON/OFF

 

 

SETUP

 

PROG

SEG

1.0.1.1.

 

0 1

0 0

PV input

 

 

calibration item

*A d J S

SETUP

PROG

SEG

0 0 0 0 0

 

0 1

0 1

Gain No. selection

 

 

* A d J S

 

 

 

SETUP

 

PROG

SEG

Current input

Zero input count

0 1

0 2

count No.

 

 

Current count No.

 

 

 

stored

After stabilized,

 

 

 

 

 

SETUP

press ENTER .

PROG

SEG

Current input

 

0 0

0 3

Span input count

count No.

 

 

Current count No.

 

 

 

stored

After stabilized,

 

 

 

 

 

SETUP

press ENTER .

Notes: *; 1. This display shows the digits shown in the previous indication.

2.If wrong key operation is made when moving from one to another item, the display is returned to the initial status of calibration mode. But, the mode is still in the calibration mode.

Figure 11-1. Calibration Flowchart (1/2)

11-4

Page 168
Image 168
Sonic Alert DCP301 user manual 11-4, Calibration Flowchart 1/2