Chapter 11. CALIBRATION

11-3 Set Up

 

PV

PV

 

CH1

CH2

34

33

28

29

Copper leads

+

+

PV

CH1

34

33

32

PV

CH2

30

29

28

 

 

Copper

 

 

leads

Millivolt sources

Millivolt sources

Gain No. 0 to 7

Gain No. 16 to 17

Decade resistance box

Decade resistance box

Gain No. 9 to 10

Gain No. 19

PV

CH1

34 33 32

PV

CH2

30

29

28

 

 

Copper

 

 

leads

PV

CH1

34

33

32

PV

CH2

30

29

28

 

 

Copper

 

 

leads

Decade resistance box Decade resistance box

Gain No. 11 to 12(0%)

Gain No. 20(0%)

PV

CH1

33 31

Copper leads

+

Signal generators 4–20mA

Decade resistance box

Decade resistance box

Gain No. 11 to 12(100%)

Gain No. 20(100%)

PV

CH2

26 27

Copper leads

-+

Millivolt sources

Gain No. 8

Gain No. 18

11-13

Page 177
Image 177
Sonic Alert DCP301 user manual 11-13