Troubleshooting Information
Checking the E5810
Checking the E5810
This section gives guidelines to check the E5810, including:
Checking Front and Rear Panel Displays
Checking E5810 Configuration Parameters
Checking Front and Rear Panel Displays
Checking LEDs
As a first step in troubleshooting the system, you should check the front panel display and LED status. In addition, you can check the rear panel Ln LED to see if there is communication from your PC to the E5810 via the LAN.
The following table shows suggested steps to troubleshoot the E5810 using the LED displays. If there is an error message on the front panel display, see “Checking Front Panel Display Error Messages”.
NOTE
The Fault LED is turned ON briefly during
LED Status | Possible Cause(s) | Corrective Action(s) |
|
|
|
All LEDs OFF | No AC power to the E5810 | Check AC power connections. |
|
|
|
Only POWER and | If FAULT is ON for more than 10 | Contact Agilent for instructions to |
FAULT ON | seconds, suspect an E5810 | return the E5810. |
| hardware failure |
|
|
|
|
POWER ON, back | Indicates no communication from | Check the LAN connection to the |
panel Ln OFF | your PC to the E5810 via the LAN. | E5810. See Chapter 1 - E5810 |
|
| Description for location of the LAN |
|
| port and the Ln LED. |
|
|
|
POWER and LAN | Indicates communication via the | Check GPIB and/or |
ON, GPIB and/or | LAN to the E5810, but not to | connections and configuration. |
connected GPIB and/or | Make sure instrument power is | |
instrument activity | instruments. | turned ON. |
|
|
|
114 | Chapter 4 |