Troubleshooting Information

Checking the E5810

Checking the E5810

This section gives guidelines to check the E5810, including:

Checking Front and Rear Panel Displays

￿Checking E5810 Configuration Parameters

Checking Front and Rear Panel Displays

Checking LEDs

As a first step in troubleshooting the system, you should check the front panel display and LED status. In addition, you can check the rear panel Ln LED to see if there is communication from your PC to the E5810 via the LAN.

The following table shows suggested steps to troubleshoot the E5810 using the LED displays. If there is an error message on the front panel display, see “Checking Front Panel Display Error Messages”.

NOTE

The Fault LED is turned ON briefly during power-on to show it is working and then turns OFF. If a hardware fault occurs, the Fault LED turns ON and remains ON for more than 10 seconds. In this case, contact Agilent Technologies.

LED Status

Possible Cause(s)

Corrective Action(s)

 

 

 

All LEDs OFF

No AC power to the E5810

Check AC power connections.

 

 

 

Only POWER and

If FAULT is ON for more than 10

Contact Agilent for instructions to

FAULT ON

seconds, suspect an E5810

return the E5810.

 

hardware failure

 

 

 

 

POWER ON, back

Indicates no communication from

Check the LAN connection to the

panel Ln OFF

your PC to the E5810 via the LAN.

E5810. See Chapter 1 - E5810

 

 

Description for location of the LAN

 

 

port and the Ln LED.

 

 

 

POWER and LAN

Indicates communication via the

Check GPIB and/or RS-232

ON, GPIB and/or

LAN to the E5810, but not to

connections and configuration.

RS-232 OFF during

connected GPIB and/or RS-232

Make sure instrument power is

instrument activity

instruments.

turned ON.

 

 

 