Using E5810 Web Access

 

Viewing and Modifying Configuration

 

 

Parameter

Description

[Default]

 

 

 

IP Address

This value is the Internet Protocol (IP) address of the E5810. The IP

[169.254.58.10]

address is a required value and is used for all IP and TCP/IP

 

communications with the E5810. The IP address is represented in dotted

 

decimal notation (for example, 154.140.222.201). This number is assigned

 

by the System Administrator. The E5810 has a default static IP address

 

of 169.254.58.10.

 

LAN traffic to the E5810 will be routed by the IP Address or by using a

 

Hostname that may be provided by a DNS name Server which is mapped

 

to this IP Address. The current E5810 IP Address may have been

 

established in one of two ways:

 

The IP address was provided by a DHCP Server during startup

 

￿(dynamic IP address).

 

The IP address shown on the Current Configuration of the E5810

 

LAN/GPIB Gateway page was used. This was set at the factory to

 

169.254.58.10. If DHCP will not be used, the IP address should be

 

set for your network.

 

 

LAN Keepalive

This value sets the LAN connect timeout in seconds. The E5810 may use

[7200 sec]

the TCP keepalive timer of the TCP/IP protocol stack to determine if a

 

client is still reachable. By specifying this configuration value, the E5810

 

turns on the keepalive timer when connecting to the client.

 

After the specified time, if there has been no activity on the connection,

 

the E5810 sends keepalive probes to the client to determine if it is still

 

alive. After a specified amount of time, the connection will be marked as

 

down (“dropped”), and the E5810 will release any resources which were

 

allocated to the associated client.

 

A value of 0 means no timeout is set. Thus, the E5810 will wait forever

 

and no keepalive probes are sent. If this value is set, it is recommended

 

that the largest value be used which still meets the application’s need for

 

unreachable client detection.

 

Smaller LAN Keepalive values will generate more keepalive probes

 

(network traffic), using more of the available network bandwidth.

 

Allowable values are 0 and greater than or equal to 600 seconds (10

 

min). The factory default setting is 7200 sec (2 hrs).

 

 

 

 