Checking Front

Panel Display Error

Messages

Troubleshooting Information

Checking the E5810

This table shows E5810 front panel display error messages and typical cause(s) for the error message.

Error Message

 

Description

 

 

Firmware Update

The firmware update failed. Verify that the IP address of the ftp server,

not successful

login, user password and filename are correct.

 

 

ERROR: NV RAM

Configuration nonvolatile data memory was corrupt. All data were

Setting defaults

restored to factory defaults, per the nonvolatile RAM (Random Access

 

Memory).

 

 

ERROR: Selftest

A power-on self-test error occurred. Errors may be viewed in the syslog

See syslog

message by typing http://<IP Address>/systemLog.htm.

 

 

ERROR: No LAN

LAN Link integrity not detected. Make sure the LAN cable is connected

connection found

and the LAN is active.

 

 

ERROR: IP in use

The E5810 tried to use an IP address that is currently being used by

<addr>

another device on the network. Try these actions:

 

 

Reboot the E5810 with the Preset button depressed to set the

 

 

E5810 back to booting with DHCP enabled.

 

 

OR

 

 

Remove other device(s) from the network and reboot the E5810

 

 

OR

 

 

Move the E5810 to a Local network, disable DHCP, and set the

 

￿

E5810 to an unused IP Address.

 

 

 

Chapter 4

115

Page 115