Troubleshooting Information

 

 

Checking PC Client Connections

 

 

 

Message

 

Description/Actions

 

 

viopen fails or I/O

The E5810 was unable to open another session or perform the

operation fails: Out

operation because it is out of resources. You may need to have a client

of resources

terminate its connection or you may need to reboot the E5810 to close

 

all client connections to reclaim resources. Rebooting will disconnect all

 

other users accessing the E5810.

 

 

I/O operation fails:

The connection to the E5810 failed because:

No connection

 

The network connection is down or the E5810 is down and a

 

 

 

 

network timeout occurred.

 

 

The E5810 was powered off and then on or an E5810 reboot was

 

 

executed, either of which terminates any open client connections.

 

 

I/O operation fails:

There are several possible causes:

Timeout

 

The client may be attempting a transaction to a non-existent or

 

 

 

 

powered-off device. Check the address used, as well as the

 

 

status of the device.

 

 

The network connection or the E5810 has gone down since the

 

 

previous I/O operation.

 

 

The I/O timeout specified is not long enough for the transaction to

 

 

complete. You may need to verify that the E5810 I/O timeout

 

 

configuration value is long enough if the client specified a timeout

 

 

value of infinity.

 

￿

The E5810 is busy performing an I/O operation for another client

 

 

or another client has the device/interface locked. See “E5810

 

 

appears to be hung” for more information.

 

 

 