Fluke 5720A service manual Cleaning the Air Filter

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 211
Image 211
Cleaning the Air Filter

Calibration and Verification 3

Cleaning the Air Filter

3-3. Cleaning the Air Filter

Caution

Damage caused by overheating may occur if the area around the fan is restricted, the intake air is too warm, or the air filter becomes clogged.

The air filter must be removed and cleaned at least every 30 days, or more frequently if the calibrator is operated in a dusty environment. The air filter is accessible from the rear panel of the calibrator.

To clean the air filter, refer to Figure 3-2 and proceed as follows:

1.Disconnect line power.

2.Remove the filter by unscrewing the knurled screw at the top of the air filter pulling the filter’s retainer downwards (it hinges at the bottom) and removing the filter element.

3.Clean the filter by washing it in soapy water. Rinse and dry it thoroughly before reinstalling.

4.Reinstall the filter and the knurled screw.

CHASSIS

 

 

GROUND

 

 

 

 

!

 

 

FUSE -F1

S2

S3

S4

F7-2.EPS

Figure 3-2. Accessing the Air Filter

3-5

Page 211
Image 211
Fluke 5720A service manual Cleaning the Air Filter