1-28. DC Voltage Specifications

5700A/5720A Series II Calibrator

Service Manual

1-28. DC Voltage Specifications

Table 1-2. 5720A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels

5720A

 

99%

99 % Confidence Level

 

Range

Resolution

 

 

Absolute Uncertainty

±5 °C from calibration temperature

For fields strengths >1 V/m but 3 V/m,

add 0.01% of range.

24 Hours

90 Days

180 Days

1 Year

±(ppm output + μV)

Relative Uncertainty

±1 °C

24 Hours

90 Days

±(ppm output + μV)

220 mV

10 nV

2.2 V

100 nV

11 V

1 μV

22 V

1 μV

220 V

10 μV

1100 V

100 μV

5 + 0.5

3.5+ 0.8

2.5+ 3

2.5+ 5

3.5+ 50

5 + 500

7 + 0.5

4 + 0.8

3 + 3

3 + 5

4 + 50

6 + 500

8 + 0.5

4.5+ 0.8

3.5+ 3

3.5+ 5

5 + 50

7 + 500

9 + 0.5

6 + 0.8

4 + 3

4 + 5

6 + 50

8 + 500

 

2

+ 0.4

2.5

+ 0.4

2

+ 0.8

 

+ 0.8

2.5

1

+ 3

 

+ 3

1.5

1

+ 5

 

+ 5

1.5

2

+ 50

 

+ 50

2.5

2.5

+ 400

 

+ 400

3

5720A

 

95%

95% Confidence Level

 

Range

Resolution

 

 

Absolute Uncertainty

±5 °C from calibration temperature

For fields strengths >1 V/m but 3 V/m,

add 0.01 % of range.

24 Hours

90 Days

180 Days

1 Year

±(ppm output + μV)

Relative Uncertainty

±1 °C

24 Hours

90 Days

±(ppm output + μV)

220 mV

10 nV

2.2 V

100 nV

11 V

1 μV

22 V

1 μV

220 V

10 μV

1100 V

100 μV

4 + 0.4

3 + 0.7

2 + 2.5

2 + 4

3 + 40

4 + 400

6 + 0.4

3.5+ 0.7

2.5+ 2.5

2.5+ 4

3.5+ 40

4.5+ 400

6.5+ 0.4

4 + 0.7

3 + 2.5

3 + 4

4 + 40

6 + 400

7.5+ 0.4

5 + 0.7

3.5+ 2.5

3.5+ 4

5 + 40

6.5+ 400

1.6

+ 0.4

2

+ 0.4

1.6

+ 0.7

 

+ 0.7

2

0.8

+ 2.5

 

+ 2.5

1.2

0.8

+ 4

 

+ 4

1.2

1.6

+ 40

 

+ 40

2

2

+ 400

 

+ 400

2.4

1-14

Page 38
Image 38