Fluke 5700A/5720A Series II Calibrator, 20. Direct Current Accuracy Test Record 5720A, Dc Dmm

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 262
Image 262
Table 3-20. Direct Current Accuracy Test Record (5720A)

5700A/5720A Series II Calibrator

Service Manual

Table 3-20. Direct Current Accuracy Test Record (5720A)

Output Current

+2A -2A

+200 mA -200 mA +20 mA -20 mA +2 mA -2 mA +200 μA -200 μA

Std

Res

Value

0.1Ω

0.1Ω

1.0Ω

1.0Ω

10Ω

10Ω

1kΩ

1kΩ

10 kΩ

10 kΩ

Characterizing dc DMM @ ±200 mV and ±2.0V

 

 

 

 

 

 

 

DC DMM

 

Column A

Column B

 

UUT Actual

 

Std Res

UUT Error

 

Setting

Rdg.

 

Error (A+B)

 

Correction

Reading

 

 

 

 

(ppm)

 

 

(ppm)

(ppm)

 

 

 

 

 

(200 mV)

(200 mV)

(200 mV)

(200 mV)

(200 mV)

(200 mV)

(2V)

(2V)

(2V)

(2V)

Test Limit,

90-day (±ppm)

118

118

57

57

38

38

39

39

49

49

3-56

Page 262
Image 262
Fluke 5700A/5720A Series II Calibrator, 20. Direct Current Accuracy Test Record 5720A, Output Current, Std Res Value