Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 269
Image 269
Table 3-27. 5700A AC Current 20 mA to 10A Accuracy Test Record

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Table 3-27. 5700A AC Current 20 mA to 10A Accuracy Test Record

2A, 200 mA, 20 mA and 10A Test Record (10A Only for 5725A)

Output

Freq

Current

 

2A

40 Hz

2A

1 kHz

2A

5 kHz

2A

10 kHz

 

 

200 mA

10 Hz

200 mA

20 Hz

200 mA

40 Hz

200 mA

1 kHz

200 mA

5 kHz

200 mA

10 kHz

 

 

20 mA

10 Hz

20 mA

20 Hz

20 mA

40 Hz

20 mA

1 kHz

20 mA

5 kHz

20 mA

10 kHz

 

 

10A (1)

40 Hz

10A (1)

1 kHz

10A (1)

5 kHz

10A (1)

10 kHz

 

 

1: If a 5725 is attached.

A40 Shunt AC-DC Diff

UUT Error

Display

Calculated

Error

Specs

±ppm

670

670

800

9100

725

400

170

170

850

2100

725

400

160

60

850

2100

417

417

888

3375

3-63

Page 269
Image 269
Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals, Output, Freq, Current