5700A/5720A Series II Calibrator

Service Manual

PREV

MENU

IF057.EPS

8.Press Oto activate the calibrator output. Then turn the output adjustment knob on the calibrator until you achieve a null on the null detector.

9.Se the rear panel CALIBRATION switch to ENABLE. Press Eon the front panel. The calibrator will now calculate a new range constant multiplier for the 220V dc range, and will store it in non-volatile memory.

10.The range calibration is now complete. Set the rear panel CALIBRATION switch to the NORMAL position, disconnect the external standards, and press rto reset the calibrator to its newly calibrated ranges.