95 % Confidence Level

Introduction and Specifications 1

AC Current Specifications

Table 1-20. 5700A Series II AC Current Specifications: 95 % Confidence Level

5700A

 

95%

95 % Confidence Level

 

Range

Resolution Frequency

Absolute Uncertainty

±5 °C from calibration temperature

For fields strengths >0.4 V/m but 3 V/m,

add 1 % of range.

Relative Uncertainty

±1 °C

Hz

24 Hours

90 Days

180 Days

1 Year

±(ppm output + nA)

24 Hours

90 Days

±(ppm output + nA)

220μA

2.2 mA

22 mA

1 nA

10 nA

100 nA

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

550 + 25

280 + 20

100 + 16

400 + 40

1300 + 80

550 + 40

280 + 35

100 + 35

400 + 400

1300 + 800

550 + 400

280 + 350

100 + 350

400 + 4000

1300 + 8000

600 + 25

310 + 20

120 + 16

500 + 40

1400 + 80

600 + 40

310 + 35

120 + 35

500 + 400

1400 + 800

600 + 400

310 + 350

120 + 350

500 + 4000

1400 + 8000

650 + 25

330 + 20

130 + 16

550 + 40

1500 + 80

650 + 40

330 + 35

130 + 35

550 + 400

1500 + 800

650 + 400

330 + 350

130 + 350

550 + 4000

1500 + 8000

700 + 25

350 + 20

140 + 16

600 + 40

1600 + 80

700 + 40

350 + 35

140 + 35

600 + 400

1600 + 800

700 + 400

350 + 350

140 + 350

600 + 4000

1600 + 8000

375 + 25

220 + 20

90 + 16

375 + 40

1200 + 80

375 + 40

220 + 35

90 + 35

375 + 400

1200 + 800

375 + 400

220 + 350

90 + 350

375 + 4000

1200 + 8000

400 + 25

250 + 20

100 + 16

400 + 40

1200 +80

400 + 40

250 + 35

100 + 35

400 + 400

1200 + 800

400 + 400

250 + 350

100 + 350

400 + 4000

1200 + 8000

Hz

±(ppm output + μA)

±(ppm output + μA)

220 mA

2.2 A

1μA

10 μA

10 - 20

20 - 40

40 - 1 k

1k - 5 k

05k - 10 k

020 - 1 k

01 k - 5 k

05 k - 10 k

550 + 4

280 + 3.5

100 + 3.5

400 + 40

1300 + 80

500 + 35

600 + 80

6500 + 160

600 + 4

310 + 3.5

120 + 3.5

500 + 40

1400 + 80

550 + 35

650 + 80

7500 + 160

650 + 4

330 + 3.5

130 + 3.5

550 + 40

1500 + 80

600 + 35

700 + 80

8000 + 1600

700 + 4

350 + 3.5

140 + 3.5

600 + 40

1600 + 80

650 + 35

750 + 80

8500 + 160

375 + 4

220 + 3.5

90 + 3.5

375 + 40

1200 + 80

500 + 35

550 + 80

6000 + 160

400 + 4

250 + 3.5

100 + 3.5

400 + 40

1200 + 80

550 + 35

650 + 80

7000 + 160

11 A

100μA

040 - 1 k

01 k - 5 k

05 k - 10 k

5725A Amplifier:

370

+ 170

400

+ 170

440

+ 170

460

+ 170

800

+ 380

850

+ 380

900

+ 380

950

+ 380

3000

+ 750

3300

+ 750

3500

+ 750

3600

+ 750

300 + 170

700 + 380

2800 + 750

330 + 170

800 + 380

3200 + 750

Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 μA and