5790A

Calibration and Verification 3

Full Verification

Note

When verifying the 5725A Amplifier at the 10A level, allow sufficient time for the A40A-10 shunt to reach thermal stability after initially applying the current.

A40 Current Shunt

Red

5790A

Input 1

792A-7004 Black

A40 Current Shunt Adapter

Pomona Cable

1368-A-18

Calibrator (UUT)

Hi

Lo

A. Alternating Current Test Setup, 22 mA to 2A Using 5790A Input 1

5790A

Input 1

792A-7004

A40 Current Shunt Adapter

A40A 10A

Pomona CableCurrent Shunt

5268-C-12

Red

Black

 

FLUKE A45-4003

 

P/N 212852

 

Cable

UHF(M) to BNC(F)

Pomona Model 1707

Adapter

UHF(F) to Banana

Kings KC-99-34

Adapter

Calibrator (UUT)

5725A (UUT)

Hi

Lo

B. Alternating Current Test Setup, 10A Using 5790A Input 1

F3-8A.EPS

Figure 3-12. Alternating Current Test Setup

3-35

Page 241
Image 241
Fluke 5720A Calibration and Verification, Full Verification, 12. Alternating Current Test Setup, 5790A, Calibrator UUT