Fluke 5700A/5720A Series II Calibrator, 37. AC V Distortion Test Summary, Load Resistors

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 276
Image 276
Table 3-37. AC V Distortion Test Summary

5700A/5720A Series II Calibrator

Service Manual

Table 3-37. AC V Distortion Test Summary

5700A/5720A Series

II Output

Load Resistors

Frequency

 

 

Measured Distortion

Max.

Distortion

2V

20V

200V

300V(1)

300V(1)

300V(1)

100Ω, 1/8W

1 kΩ, 1/2W

10 kΩ, 5W

15kΩ, 5W 15 kΩ, 5W 15 kΩ, 5W

10 Hz, 20 Hz

1 kHz, 20 kHz, 50 kHz,

100 kHz, 200 kHz, 500 kHz

10 Hz, 20 Hz

20 kHz, 100 kHz

200 kHz, 500 kHz

10 Hz, 20 Hz

50 kHz, 100 kHz

40Hz

50 kHz

70 kHz

0.054%

0.044%

0.355%

0.0535%

0.0385%

0.304%

0.055%

0.1065%

0.1%

0.3%

0.4%

(1)The 5700A/5720A Series II maximum volt-Hertz product is (2.2 x 107) The 300V level assumes that a Fluke 5725A Amplifier is attached.

Table 3-38. Test Record for Flatness Check of the AC 2 mV Range

Frequency

10Hz

20Hz

40Hz

1kHz

20kHz

50kHz

100kHz

300kHz

500kHz

1MHz

UUT Error Display Reading

Limits

±0.26%

±0.26%

±0.29%

±0.49%

±0.87%

±1.83%

3-70

Page 276