Auxiliary Amplifier Specifications

Introduction and Specifications 1

Auxiliary Amplifier Specifications

1-35. Auxiliary Amplifier Specifications

For complete specifications, see the 5205A and 5220A Operators Manuals.

5205A (220V - 1100 V ac, 0 V - 1100 V dc)

Overshoot: < 10 %

Distortion (bandwidth 10 Hz - 1 MHz):

10 Hz - 20 kHz

20 kHz - 50 kHz

50 kHz - 100 kHz

Frequency

(Hz)

dc 10 - 40 40 - 20 k

20 k - 50 k

50 k - 100 k

0.07%

0.2%

0.25%

90 Day Accuracy at 23° ±5 °C

±(% output + % range)

0.05+ 0.005

0.15+ 0.005

0.04+ 0.004

0.08+ 0.006

0.1+ 0.01

Temperature Coefficient for 0°-18 °C and 28°-50 °C
±(ppm output + ppm range)/ °C

15 + 3

45 + 3

15 + 3

50 + 10

70 + 20

5220A (AC Current, 180-day specifications):
Accuracy:

 

20 Hz - 1 kHz

0.07% + 1 mA

1 kHz - 5 kHz

(0.07% + 1mA) x frequency in kHz

Temperature Coefficient (0° - 18 °C and 28° - 50 °C): (0.003 % + 100 μA)/°C
Distortion (bandwidth 300 kHz):

20 Hz - 1 kHz

0.1% + 1 mA

1 kHz - 5 kHz

(0.1% + 1 mA) x frequency in kHz

Note: 5700A/5720A combined with 5220A is not specified for inductive loads.