Use Table 3-23 (5720A) or Table 3-24 (5700A) for the test records.

5700A/5720A Series II Calibrator

Service Manual

Use Table 3-23 (5720A) or Table 3-24 (5700A) for the test records.

1.Place the 5790A on top of the Calibrator and connect the equipment as shown in Figure 3-11.

Note

The point of measurement is at the end of the cable and adapter that connects to the 5790A. Other cable lengths and adapters will yield different results at high frequencies.

2.On the 5790A push UTIL MENUS button and then the MEAS CONTROL softkey. Set the digital filter mode to FAST and the restart to MEDIUM. Push the DONE soft key twice to return to the measurement display.

3.Set the 5790A to 2 mV at 1 kHz. Adjust the Calibrator using the output adjustment knob for a reading of 2.0000 mV ±1 count on the 5790A. Record the Calibrator error display reading in the 90 Day column in Table 3-23 (5720A) or 3-24 (5700A) as appropriate for the verification interval.

4.Verify that the result is within the test limits.

5.Repeat the previous steps for the 2 mV output on all remaining frequencies on the test record.

6.Proceed to the remaining output levels and frequencies list in Table 3-23 or 3-24 and repeat steps 3 through 5 using the appropriate output level in each step, and the adjustment tolerance in Table 3-8.

BNC to

Banana Adapter

5790A Side View

HI

Input 2

12 Inch Coax Cable Manual background

BNC to

LO

Banana Adapter

Calibrator Side View

HI

Output

LO

F3-6.EPS

Figure 3-11. AC Voltage Test Setup