Fluke service manual 5700A/5720A Series II Calibrator, Service Manual

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 40
Image 40
Overshoot: <5 %

5700A/5720A Series II Calibrator

Service Manual

Table 1-4. DC Voltage Secondary Performance Specifications and Operating Characteristics

Temperature Coefficient

Adder [Note 2]

Noise

Range

220mV

2.2 V

11V

22V

220V

1100 V

Stability [Note 1]

±1 °C

24 Hours

±(ppm output + μV)

0.3+ 0.3

0.3+ 1

0.3+ 2.5

0.4+ 5

0.5+ 40

0.5+ 200

10 °-40 °C

0 °-10 °C

and 40°-50 °C

±(ppm output + μV)/°C

0.4

+ 0.1

1.5

+ 0.5

0.3

+ 0.1

1.5

+ 2

0.15

+ 0.2

1

+ 1.5

0.2

+ 0.4

1.5

+ 3

0.3

+ 5

1.5

+ 40

0.5

+ 10

3

+ 200

Linearity

Bandwidth

± 1 °C

0.1-10 Hz

 

 

 

pk-pk

 

 

±(ppm output + μV)

1

+ 0.2

0.15

+ 0.1

1

+ 0.6

0.15

+ 0.4

0.3

+ 2

0.15

+ 2

0.3

+ 4

0.15

+ 4

1

+ 40

0.15

+ 60

1

+ 200

0.15

+ 300

Bandwidth

10 Hz-10 kHz

RMS

μV

5

15

50

50

150

500

Notes:

1.Stability specifications are included in the Absolute Uncertainty values in the primary specification tables.

2.Temperature coefficient is an adder to uncertainty specifications that does not apply unless operating more than ±5 °C from calibration temperature.

Minimum output: 0 V for all ranges, except 100 V for 1100 V range

Maximum load: 50 mA for 2.2 V through 220 V ranges; 20 mA for 1100 V range; 50 Ω output impedance on 220 mV range; all ranges <1000 pF, >25 Ω

Load regulation: <(0.2 ppm of output + 0.1ppm of range), full load to no load

Line regulation: <0.1 ppm change, ± 10 % of selected nominal line

Settling time: 3 seconds to full accuracy; + 1 second for range or polarity change; + 1 second for 1100 V range

Overshoot: <5 %

Common mode rejection: 140 dB, DC to 400 Hz

Remote sensing: Available 0 V to ±1100 V, on 2.2 V through 1100 V ranges

1-16

Page 40
Image 40
Fluke 5700A/5720A Series II Calibrator, Service Manual, Overshoot 5 % Common mode rejection 140 dB, DC to 400 Hz