RESISTANCE STANDARD

5700A/5720A Series II Calibrator

Service Manual

RESISTANCE STANDARD

CALIBRATOR

 

 

CURRENT

SENSE

OUTPUT

SENSE

VΩA

VΩ

 

 

HI

HI

HI

HI WIDEBAND

LO

LO

 

 

 

LO

LO

 

 

 

 

HI

 

 

CHASSIS

AUX

GUARD GROUND

 

GROUND

CURRENT

 

F7-4.EPS

Figure 3-4. 742A-1 and 742A-10k External Calibration Connections

10.To reject the changes, return to the calibration menu shown in step 2 by pressing P. Otherwise, press the softkey under “Proceed” to accept and save the changes, and to open the display shown below, letting you continue with calibration.

PREV

MENU

IF3-12.EPS

3-12

Page 218
Image 218
Fluke 5720A Resistance Standard, Calibrator, 4. 742A-1 and 742A-10k External Calibration Connections, F7-4.EPS, IF3-12.EPS