Image 216
IF3-5.EPS

5700A/5720A Series II Calibrator

Service Manual

6.Enter the true value of the 732B 10V output. The true value is the value printed on the calibration sticker, plus or minus, as well as drift that has occurred since calibration. That drift can be estimated provided control charts have been maintained for the 732B.

If the entered value is not between 9V and 11V, and error message appears, which lets you start over from this point with a calibrated 732B. Press E, and the display shows the following:

PREV

MENU

IF3-5.EPS

When the calibrator’s 6.5V and 13V references have been characterized, the display shows the following message, which lets you accept or reject the changes that are about to be made to the calibration constants.

PREV

MENU

IF3-6.EPS

7.To reject the changes, return to the calibration menu shown in step 2 by pressing P. Otherwise, press the softkey under “Proceed” to accept and save the changes, and to open the display shown below, letting you continue with calibration.

PREV

MENU

IF3-7.EPS