Press Pto exit the calibration menus.

Calibration and Verification 3

Performing a Calibration Check

PREV

MENU

IF3-3.EPS

3.Enter the ambient temperature; then press E. As the calibration check proceeds, the display indicates the current process of the calibration check. When the check is complete, the largest shift that is detected appears on the display.

PREV

MENU

IF3-17.EPS

4.You can now print a list of the proposed shifts, or quit without creating a list.

To print the report, connect a printer and set up the serial interface as described in Chapter 6 of the 5700A/5720A Operators Manual. Press the softkey under “Print Output Changes.” To return to normal operation without printing, press the softkey under “DONE with check.”

Press Pto exit the calibration menus.

3-21